User guide
4.4 Basic settings
732 IC Detector / 733 IC Separation Center
91
>CONFIG/print meas.value
time interval 1.0 s
0.4…99999 s
Time interval for measured value printout
This inquiry appears only with "print crit. = time".
Time interval in seconds between printout of the indi-
vidual measured values.
>CONFIG/print meas.value
stop time off min
1…999 min, off
Stop time for measured value printout
This inquiry appears only with "print crit. = time".
Time to stop of the measured value printout.
off No time restriction
(press <CLEAR>).
>CONFIG/print meas.value
date&time: off
on, off
Printout of date and time
This inquiry appears only with "print crit. = immed." or "print
crit. = time".
on Date and time are printed out for every
measured value.
off Date and time are not printed out.
>CONFIG
>CONFIG/auxiliaries
General instrument settings
>CONFIG/auxiliaries
run number 0
0…999, off
Run number
The run number is incremented by +1 for each new
determination as follows:
without program Increase with every "INJECT A".
with program Increase with every new start and
with every "return" in cycle pro-
grams.
The parameter value has the following meaning:
0…999 Start point for numbering.
off No numbering
(press <CLEAR>).
>CONFIG/auxiliaries
number of cycles 1
1…999
Number of cycles
This parameter is identical to the parameter of the same
name under the <PROGRAM> key (see section 4.7.1)
and determines how many times cycle programs should
be run. As access to the <PROGRAM> key is blocked
while a program is running, the number of cycles can
subsequently be changed here.