User guide
4 Operation
732 IC Detector / 733 IC Separation Center
86
>SETUP/input assign
ZERO OFF 9
1…15
Switch off auto-zero
Equivalent to pressing the <ZERO OFF> key.
>SETUP/input assign
PLOT 10
1…15
Plot output
Equivalent to pressing the <PLOT> key.
>SETUP/input assign
polarity 11
1…15
Switch the polarity
The polarity of the signal at the analog output is
switched.
>SETUP/input assign
SELECT 13
1…15
Select
Equivalent to pressing the <SELECT> key.
>SETUP/input assign
QUIT 14
1…15
Quit
Equivalent to pressing the <QUIT> key.
>SETUP/input assign
ENTER 15
1…15
Enter
Equivalent to pressing the <ENTER> key.
>SETUP
>SETUP/graphics
General graphics parameters
General parameters for the printout of graphics plots on an
external printer.
>SETUP/graphics
grid: off
on, off
Grid lines for graphics plot
on Dotted grid lines are plotted on the graphics
printout.
Off No grid lines are plotted on the graphics
printout.
>SETUP/graphics
frame: off
on, off
Frame for graphics plot
on A frame is plotted on the graphics printout.
Off Only the y and x axes are plotted on the
graphics printout, but no frame.
>SETUP/graphics
width 0.8
0.4…1.0
Relative width of the graphics printout
The width of the graphics printout must be matched to
the attached printer.