User guide
4.3 Displays in the standby mode
732 IC Detector / 733 IC Separation Center
83
stopped: YY-MM-DD HH:MM
The 709 IC Pump was stopped at the time
indicated owing to violation of the shutoff
limit "Pmin" or "Pmax". To restart a pump
stopped in this manner, the pump keys
<EXT.> and <R/S> must be pressed in
succession and then <EXT.> pressed
again (see also section 4.5.1).
pump not responding
This message appears when
• the 709 IC Pump is not connected to the
732 IC Detector,
• the 709 IC Pump is not switched on,
• the remote control of the 709 IC Pump is
not switched on.
+0.004 µµS/cm 13.5 min
prog.type cycle no. XXX
Status messages for program type
These status messages appear only with
"PROGRAM/parameters/status = active".
prog.type cycle no. XXX
Display of the program type "cycle" and the
cycle number (see section 4.7.1).
prog.type inject
Display of the program type "inject" (see
section 4.7.1).
prog.type remote
Display of the program type "remote" (see
section 4.7.1).
+0.004 µµS/cm 13.5 min
next step ready
Status messages for program step
These status messages appear only with
"PROGRAM/parameters/status = active".
next step ready
The program is ready for a new start (see
section 4.7.1).
next step XXX.X min
Display of the time when the next program
step will be executed.
next step ---
The program has been set to "inactive", a
new start is not possible.
next step ì
The program continues to run indefinitely (no
"end" command available).