User guide
4 Operation
732 IC Detector / 733 IC Separation Center
76
Auxiliary function keys of 732 IC Detector
Key Standby mode Edit mode
SELECT
Select status message
• Selection of the status messages for
full-scale range, absolute conductiv-
ity, 709 IC Pump, program, method,
date and time.
Select preset parameter values
• Selection from preset parameter
values for parameters which are
marked by a colon ":".
CLEAR
No function Delete parameter values
• Deletion of the displayed parameter
values and resetting to default values
or "0".
QUIT
Acknowledge error messages
• Acknowledgment of the error mes-
sages displayed on the first line.
or
abort printing operations
• Abort printing operations started with
<PLOT>, <REPORT> or <PRINT>.
Abort entry
• Abort the parameter entry (parameter
reset to original value).
• Exit from rolling inquiries and selec-
tion of the next higher menu level or
the standby mode.
ENTER
No function Confirmation of parameters
• Confirmation of existing parameter
values or those just entered.
• Selection of the next menu line.
Operating keys of 733 IC Separation Center
Key Injection valve Suppressor module
FILL
Switching to "FILL"
• Switching of injection valve A to "FILL"
position.
No function
INJECT
Switching to "INJECT"
• Switching of injection valve A to
"INJECT" position.
No function
STEP
FILL
Switching to "FILL"
• Switching of injection valve B to
"FILL" position.
Rotate suppressor module
• Switch suppressor module to next
position.
INJECT
Switching to "INJECT"
• Switching of injection valve B to
"INJECT" position.
No function