User guide
4.2 Fundamentals of the operation
732 IC Detector / 733 IC Separation Center
73
Main function keys of 732 IC Detector
Key Standby mode Edit mode
ZERO
Trigger auto-zero
• Automatic zeroing of the current
conductivity (the green LED in the
key lights up).
Trigger auto-zero
• Automatic zeroing of the current
conductivity (the green LED in the
key lights up).
PARAM
Open parameter menu
• Opens the main menu for setting the
parameters for the conductivity de-
tector, analog output, graphics plot
and 709 IC Pump.
• All settings which are set in the
parameter menu can be stored to-
gether with the program as a
method.
Select submenu or parameters
• Selection of the next submenu in the
main menu.
• Selection of the next parameter in the
submenu.
SCALE
FULL
Set full-scale range
• Direct selection of the full-scale range
(value flashes).
• Selection of the value with the <ç>
and <è> keys.
Return to standby mode
• Acceptance of the displayed value for
the full-scale range and return to the
standby mode.
R/S
PROG
Start/stop program
• LED in key dark:
Program inactive; No function
(program start not possible until pro-
gram status = "active").
• LED in key lit up:
Program active; Program start with
program type "cycle".
• LED in key flashes:
Program running; Program stop
with all program types.
No function