3.4 Calibration
732 IC Detector / 733 IC Separation Center
61
+0.000 µµS/cm 0.0 min
full scale: 4.00 µµS/cm
As soon as injection valve A is in the "INJECT" position, the
current time is reset to "0.0 min". This marks the start time
of the chromatogram.
The ion chromatogram of the standard is plotted in the next
20 min or so.
Fig. 19 shows an example of a calibration with 5 ppm Cl
,
10 ppm NO
3
and 10 ppm SO
4
2–
.
7
.
00e+005
6
.
00e+005
5
.
00e+005
4
.
00e+005
3
.
00e+005
2
.
00e+005
0
.
00e+005
1
.
00e+005
0 2 4 6 8 10 12 14 16 18
FULL REPORT
Ret Time Component Concentr. Area Height
(Min) Name ppm (uV*Sec) (uV)
-------- --------------- ----------- ---------- ----------
1.758 Chloride 5.000000 2791676.00 416181.718
3.164 Nitrate 10.000000 2405831.75 202204.046
10.575 Sulphate 10.000000 4652037.50 146991.609
Fig. 19: Ion chromatogram of the calibration
SO
4
2–
NO
3
Cl
Standard