User guide

2.7 Precolumns
732 IC Detector / 733 IC Separation Center
31
2.7.4 IC anion precolumn SUPERSEP
The 6.1009.010 IC Anion Precolumn SUPERSEP has two connections
for steel or PEEK capillaries and is installed as follows:
1 Connect precolumn
Remove end caps from the precolumn.
Fit connectors to column connection capillary 6767 mounted on
the injection valve (see section 2.5).
Screw precolumn to column connection capillary 6767.
Cut a piece from the 6.2620.020 Steel capillary or the
6.1831.010 PEEK capillary as short as possible and fit with
connectors (see section 2.5).
Fasten the prepared capillary to the other end of the precol-
umn.
2 Rinse precolumn
Place a beaker beneath the outlet capillary of the precolumn.
Switch on 709 IC Pump and rinse precolumn with eluent for
ca. 10 min.
Switch off IC pump.