User guide

2 Installation
732 IC Detector / 733 IC Separation Center
26
2 Installation of the capillary in the IC Separation
Center
Unscrew the four knurled screws 3737 of the bottom rear panel
3939 of the 733 IC Separation Center and remove rear panel
(see Fig. 5).
Insert steel capillary 7171 from the back into the inner com-
partment of the IC Separation Center.
Install rear panel 3939 so that the capillary is positioned in
opening 4040 „Inlet A“ or 4444 „Inlet B“ and screw on with the four
knurled screws 3737.
3 Connection to injection valve
Connect steel capillary 7171 to pulsation dampener 6666 (see
section 2.6.2). Using another steel capillary 7171, connect pul-
sation dampener 6666 to connection "5" of injection valve 6868.
4 Mount column connection capillary
Connect column connection capillary 6767 (ca. 20 cm of
6.1831.010 PEEK capillary) to connection "4" of injection valve
6868.
709
5656 7070
733
71716969 6363 7171 6666 6767 6868
Fig. 14: Connection to injection valve with steel capillaries
5656 Inlet capillary
of Filter unit Manufit 6969
6868 Injection valve
6363 Outlet capillary
of Filter unit Manufit 6969
6969 Filter unit Manufit (6.2821.000)
6666 Pulsation dampener (6.2620.150) 7070 Coupling (6.2620.060)
6767 Column connection capillary
6.2620.020 Steel capillary
7171 Steel capillary (6.2620.020)