User guide
2.6 Connection of 709 IC Pump
732 IC Detector / 733 IC Separation Center
23
2.6.4 Filter unit Manufit
The optional 6.2821.000 Filter unit Manufit (see section 7.3.1) is in-
stalled between the 709 IC Pump and the injection valve at the 733 IC
Separation Center. The filter unit serves to avoid contamination of the
piston seals by abrasive particles and can be used in the pressure
range 0…50 MPa (0…500 bar) for non metal-free systems together
with steel capillaries. It is installed as follows (see Fig. 12):
1 Prepare Manufit housing
• Insert outlet capillary 6363 with steel mesh holding end 6161 into
Manufit housing 6262.
• Insert the 4 steel meshes 6060 provided into the steel mesh
holding end 6161.
• Press the PTFE gasket 5959 into the steel mesh holding end
6161.
2 Prepare Manufit pressure screw
• Insert inlet capillary 5656 with counterpart end 5858 into Manufit
pressure screw 5757.
3 Assembly
• Fit the two capillary end pieces 5858 and 6161 together.
• Screw Manufit pressure screw 5757 and Manufit housing 6262
firmly together.
To replace contaminated steel meshes, proceed in the reverse order.
IC Separation
Center 733
IC Pump
709
5656
5757
5858
5959
6060
6161
6262
6363
Fig. 12: 6.2821.000 Filter unit Manufit
5656 Inlet capillary
5757 Manufit pressure screw
5858 Counterpart end
5959 PTFE Gasket (6.2821.010)
6060 Steel meshes (6.2821.020)
6161 Steel mesh holding end
6262 Manufit housing
6363 Outlet capillary