User guide
6 Interfaces
732 IC Detector / 733 IC Separation Center
188
6.5 Valve interfaces
The two valve interfaces 3838 and 4646 of the 733 IC Separation Center can
be used for the attachment of external devices (e.g. 750 Autosampler,
«IC Metrodata for Win95») which can be employed to control the injec-
tion valves or the suppressor module (see section 2.3). The two valve
interfaces for the injection valves A and B (or the suppressor module)
are identical and have the following connections
733 IC Sep. Center Connection Function
Inputs
0V
Ground Ground
+5V
10kΩ
Fill
Valve →→ "FILL"
Valve is switched to the position "FILL"
(or suppressor module is switched further)
H
L
t
P
> 2 ms
+5V
10kΩ
Inject
Valve →→ "INJECT"
Valve is switched to the position "INJECT"
H
L
t
P
> 2 ms
Outputs
COM
RUN
Position "Fill"
A pulse is outputted when the valve is switched to
the position "FILL".
H
L
t
P
≅≅ 250 ms
COM
RUN
Integrator Start
A pulse is outputted when the valve is switched to
the position "INJECT".
H
L
t
P
≅≅ 250 ms
No liability whatsoever will be accepted for damage arising from the improper connection of devices.