User guide

6 Interfaces
732 IC Detector / 733 IC Separation Center
172
Object Meaning Entry range/Selection
&Setup Operating mode
³
Ã.IdReport
³
³
³
³
³
³
³
³
³
³
³
³
³
Send identification before report
Elements of the message:
®
Space (dec 32), ', report identification
¯
" 'co" config
" 'ev" event
" 'me" method
" 'ml" method list
" 'cp" parameter
" 'pr" program
" 'pu" pump
" 'mp" measured value output with
<PRINT>
on, off
³
Ã.Keycode
³
³
³
³
³
³
³
³
³
³
³
³
³
³
³
Send key code of pressed keys
Elements of the message:
®
Space (dec 32), # or ù, two-place code
¯
0 9 (METHOD) 12 CLEAR
1 8 (PUMP R/S) 13 SELECT
2 7 (CONFIG) 14 3 (PRINT)
3 ZERO 15 2 (REPORT)
4 PARAM 16 1 (PLOT)
6 6 (PROGRAM) 18 ENTER
7 5 (EVENT) 19 QUIT
8 4 (ZERO OFF) 20 –/+ ()
9 PROG R/S 21 . ()
10 FULL SCALE 22 0 (MARK)
on, off
³
Ã.Tree Definition of the response to $Q
³ Ã.Short
³ ³
³ ³
Path names are sent with just the required number
of characters (boldface characters)
on, off
³ À.ChangedOnly
³
Only path names and their values which have
been edited once are sent.
on, off
³
Ã.Trace
³
³
³
Send path and value on changes
Element of the change message:
®
Space (dec 32), path, ”value”
¯
on, off
³
Ã.Lock Lock functions
³ Ã.Keyboard Lock all keys on, off
³ Ã.Config Locking of <CONFIG> key on, off
³ Ã.Event Locking of <EVENT> key on, off
³ Ã.FullScale Locking of <FULL SCALE> key on, off
³ Ã.Mark Locking of <MARK> key on, off
³ Ã.Method Locking of <METHOD> key on, off
³ Ã.Param Locking of <PARAM> key on, off
³ Ã.Plot Locking of <PLOT> key on, off
³ Ã.Print Locking of <PRINT> key on, off
³ Ã.ProgRS Locking of <PROG R/S> key on, off
³ Ã.Program Locking of <PROGRAM> key on, off
³ Ã.PumpRS Locking of <PUMP R/S> key on, off
³ Ã.Report Locking of <REPORT> key on, off
³ Ã.Select Locking of <SELECT> key on, off
³ Ã.Zero Locking of <ZERO> key on, off
³ Ã.Zerooff Locking of <ZERO OFF> key on, off