User guide

6.1 RS232 interfaces
732 IC Detector / 733 IC Separation Center
171
Object Meaning Entry range/Selection
³ Ã.Change
³ ³
³ ³
³ ³
Change in the status of the input lines since the
last ”clear”, in byte form (1 = change,
0 = no change, see above)
read only
³ À.Clear Reset ”change” byte to 0 $G
³
³
³
Ã.Outputs Remote outputs
³ Ã.State
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
Status of the output lines of the two remote
interfaces "733 IC Separation Center" and
"Remote" in byte form: 1=on, low, active;
0=off, high, inactive
"733 IC Sep. Center " "Remote"
ÚÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ¿ ÚÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ¿
n 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
Pin 5 18 4 17 3 16 1 2 5 18 4 17 3 16 1 2
Output 8 7 6 1 2 3 4 5
Status
n
n
=
=
2 2
0
14
15
Example:
0 0 1 0 0 0 0 0 1 0 0 0 1 0 0 1
Status = 2
2
+ 2
8
+ 2
12
– 2
15
=
= 4 + 256 + 4096 – 32768 = –28412
read only
³ Ã.Change
³ ³
³ ³
³ ³
Change in the status of the output lines since the
last ”clear”, in byte form (1 = change,
0 = no change, see above)
read only
³ À.Clear Reset ”change” byte to 0 $G
³
Ã.MeasValue Measured values
³ Ã.Conductivity Current measured value in S/cm read only
³ Ã.Background Absolute conductivity in S/cm read only
³ À.RunTime Run time in min read only
³
Ã.Pump709 709 IC Pump
³ Ã.State Status: ready, running, stopped, unknown read only
³ Ã.Pactual Current pressure in MPa read only
³ À.Error
³
³
Error status: no error, stopped:
YY-MM-DD XX:MM (stopped with date)
read only
Ã.Method Method in working memory
³ Ã.Name Method name read only
³ Ã.Status Status: original, modified read only
³ À.Index Method index (110) read only
³
À.Display Display
Ã.Select
³
³
Status display: full scale, absolute, prog.type,
prog.step, method, date
read only
Ã.L1 Display in LCD line 1 read only
Ã.L2 Display in LCD line 2 read only
À.LED Status of the LED displays in byte form:
1=on, 0=off
n 0 1 2 3
LED OVERLOAD THERMOSTAT PROG R/S ZERO
Status
n
n
=
=
2
0
3
Example: 0 1 1 1
Status = 2
1
+ 2
2
+ 2
3
= 2 + 4 + 8 = 14