User guide
6.1 RS232 interfaces
732 IC Detector / 733 IC Separation Center
167
Object Meaning Entry range/Selection
Ã.Aux General instrument settings
³ Ã.RunNo
³ ³
Run number (automatically incremented by 1
after every determination)
0…999, off
³ Ã.Cycles Number of cycles in loop programs 1…999
³ Ã.CEvent Settings for events
³ ³ À.CEenable Perform events during running program on, off
³ Ã.Set Setting of date and time $G, $S
³ ³ Ã.Date Date YYYY–MM–DD
³ ³ À.Time Time HH:MM:SS
³ Ã.Dialog
³ ³
Dialog language english, deutsch, francais,
español
³ Ã.DevLabel Device label 8 ASCII characters
³ Ã.Prog Number of program version read only
³ À.Beeper Beeper
³ Ã.Status Status: on, error only, off on, only error, off
³ À.Repeat Repeat time for error message 5…60…999 s, off
³
Ã.RSset RS232 interface
³ Ã.Baud
³ ³
Baud rate in bit/s 9600, 4800, 2400, 1200, 600,
300
³ Ã.DataBit Data bits 7, 8
³ Ã.StopBit Stop bits 1, 2
³ Ã.Parity Parity: none, odd, even none, odd, even
³ À.Handsh
³
Handshake: hardware simple, hardware full,
software character, software line, none
HWs, HWf, SWchar, SWline,
none
³
Ã.RSSet709 2
nd
RS232 interface "709 IC Pump"
³ Ã.Status Connection in operation without 709 IC Pump printer, off
³ Ã.709 Connected: 709 IC Pump
³ ³ À.Baud
³ ³
Baud rate in bit/s 9600, 4800, 2400, 1200, 600,
300
³ À.Printer Connected: printer
³ Ã.Baud
³ ³
Baud rate in bit/s 9600, 4800, 2400, 1200, 600,
300
³ Ã.DataBit Data bits 7, 8
³ Ã.StopBit Stop bits 1, 2
³ Ã.Parity Parity: none, odd, even none, odd, even
³ À.Handsh
³
Handshake: hardware simple, hardware full,
software character, software line, none
HWs, HWf, SWchar, SWline,
none
³
À.SepCenter 733 IC Separation Center
Ã.Config Display of the configuration read only
Ã.Control Control: without restriction, via 732 only no restriction, 732 only
Ã.AValve Injection valve A
³ Ã.StatusA Display of valve setting: inject, fill read only
³ À.TriggerA Switching the valve inject, fill
Ã.BValve Injection valve B
³ Ã.StatusB Display of the valve setting: inject, fill read only
³ À.TriggerB Switching the valve inject, fill
À.Suppressor Suppressor module
Ã.Autostep Automatic triggering of "step" fill, inject, off
Ã.Status
³
³
Display of the suppressor setting:
in position or undefined
read only
À.Trigger Switching the suppressor ---, step