User guide
6 Interfaces
732 IC Detector / 733 IC Separation Center
166
6.1.6 Remote control commands
The remote control tree can be divided into the following main
branches:
& Root
à Config Instrument configuration
à Methods Methods
à Parameter Parameters
à Program Program
à SelectEvent Events
à Zero On/off switching of auto-zero
à Marker Triggering marking signal
à Plot Start/stop graphics plot
à Print Start/stop measured value printout
à Info Instrument information
à Setup Settings of operating mode
à Assembly Data of the assemblies
À Diagnosis Diagnostic tests
The following table lists all objects of the remote control tree. For the
unambiguous designation of the objects, the boldface characters suf-
fice. The meaning of the individual object is described here only in brief,
for more detailed information please refer to section 4. The default val-
ues of the objects are printed in boldface.
Object Meaning Entry range/Selection
&Config Instrument configuration
³
Ã.Detector Conductivity detector
³ Ã.Thermostat Operating temperature of conductivity cell 25, 30, 35, 40, 45°C, off
³ Ã.ZeroUnit Unit for display of auto-zero value µS/cm, %fs, mV
³ À.CellConst Cell constant of conductivity cell 13.0…16.7…21.0 /cm
³
Ã.Printer External printer
³ Ã.Id1 1
st
line of printout header 18 ASCII characters
³ Ã.Id2 2
nd
line of printout header 18 ASCII characters
³ Ã.PrintHead Output of printout header: once, always, never once, always, off
³ Ã.DateTime Output of date and time in printout header on, off
³ À.SendTo
³
Selection of printer driver IBM, Epson, Seiko, Citizen,
HP
³
Ã.PrintMeasVal
³ ³
Measured value printout
(function same as <PRINT> key)
$G, $S
³ Ã.PrintCrit
³ ³
Criterion for measured value printout: immediately
(single value), time (continuous output), off
immed., time, off
³ Ã.Time Timed output of measured values
³ ³ Ã.Interval Time interval for output of measured values 0.4…1.0…99999 s
³ ³ À.StopTime Stop time (off = infinite) 1…999 min, off
³ À.DateTime Output of date and time of measured value on, off