User guide

6 Interfaces
732 IC Detector / 733 IC Separation Center
164
6.1.4 Status messages
To ensure appropriate control of an external control unit is possible, it
must also be possible to request status conditions which provide in-
formation on the status of the 732 IC Detector. The output of a status
message is triggered by the trigger '$D'. Status messages comprise
the global status '$R' and the detailed status messages listed below,
which can also appear in combination, e.g. '$R.Cond.ProgWait'. If a
fault appears, the error message is also appended to the status mes-
sage, e.g. '$R.Cond.ZeroOK;E42'.
$R Ready: The 732 IC Detector is ready to measure.
.Cond Basic status of the conductivity measurement
.Zero Auto-zero function active, value not yet valid
.ZeroOK Auto-zero function active, value valid
.Overflow The measured value is 5080% outside the full-scale range
.Background The measured value is more than 80% outside the full-scale
range
.ProgWait Program active (can be started)
.ProgGo Program running
.ProgEnd Program ended
.ProgTest Program test mode active
.Input Change in the signal of an input line
.Output Change in the signal of an output line
.Diagnosis Instrument in diagnostic status