User guide
5 Notes – Maintenance – Faults
732 IC Detector / 733 IC Separation Center
158
• Press <ENTER> key to open the following diagnostic menu:
>diag/EEPROM initial.
block: USER
üü üü üü üü üü üü
>diag/EEPROM initial.
block: CONFIG
üü
>diag/EEPROM initial.
block: INFO
üü
>diag/EEPROM initial.
block: SETUP
üü
>diag/EEPROM initial.
block: ASSEMBLY
üü
>diag/EEPROM initial.
block: EVENT
üü
>diag/EEPROM initial.
block: METHOD
üü
>diag/EEPROM initial.
block: ALL
üü üü üü üü üü üü üü üü
1)
Set config parameters with default values. Ã
Set info parameters with default values. Ã
Set setup parameters with default values. Ã
Set assembly parameters with default values. Ã
Set event parameters with default values. Ã
Delete method memory. Ã
Delete device number. Ã
Set instrument adjustment with default values. Ã
Pressing the <SELECT> key selects the submenus in turn. The individual initializa-
tion alternatives are accessed using the <ENTER> key, exit is with the <QUIT>
key.
The table shows which parts of the data memory are affected by the corresponding
initialization alternatives.
• If need be, press the <SELECT> key repeatedly until the desired
initialization alternative appears (such as “USER“)
>diag/EEPROM initial.
block: USER
• <ENTER>
After the initialization has run automatically, the instrument quits the diagnostic menu
and runs a power on reset.
1)
Care: This initialization alternative is reserved for the service engineer!