User guide

5 Notes – Maintenance – Faults
732 IC Detector / 733 IC Separation Center
148
5.3.2 Malfunctions and their rectification
If difficulties appear with the IC system during analyses, their causes
are best investigated in the order separating column pump
eluent 732/733 IC System. Several of the malfunctions which may
appear are listed in the following table with details of possible causes
and countermeasures.
Malfunction Cause Rectification
Baseline with high
noise level,
pulsation
Contaminated pump
values
Faulty piston seals
Quality of the pump
does not suffice for the
selected sensitivity
Clean the valves (see
section 5.1.2)
Replace the piston seals
(see section 5.1.2)
Use pulsation dampener,
use more powerful pump
or lower the sensitivity
Drift of the
baseline
Thermal equilibrium not
yet reached
Leak in system
Evaporation of organic
solvent in eluent
Condition system with
heating switched on
Check connections and
make leakproof
Ensure better closure of
eluent supply vessel
Considerable
pressure drop
Leak in system Check connections and
make leakproof
Considerable
pressure rise
Contamination of the
filter in the 6.2821.100
Filter unit PEEK
Contamination of the
filter in the 6.2821.000
Filter unit Manufit
Contamination of the
column inlet filter
Change of column
packing by injection of
contaminated samples
Replace the 6.2821.110
Filter (see section 2.6.3)
Clean or replace
6.2821.020 Steel mesh(es)
Clean or replace
6.2821.020 Steel mesh(es)
Regenerate the column
(see section 5.1.1) or re-
place column
Note:
Samples should always
be microfiltered.
Chromatograms
with poor resolu-
tion, change in the
retention times
Deterioration in separa-
tion efficiency of the IC
column
Regenerate the column
(see section 5.1.1) or re-
place column
Extreme peak
broadening,
splitting (double
peaks)
Dead volume at the
column ends
Fill column with glass
beads ( 100 µm) or
replace column
No feed of re-
generation or
rinsing solution
for suppressor
Contamination of the
filter in the 6.2821.100
Filter unit PEEK
Counterpressure too
high in suppressor
module
Replace the 6.2821.110
Filter (see section 2.6.3)
Clean or replace suppres-
sor module (see section
5.2.65.2.8)