3 Clean input and output leads
Connect each of the 6 capillary tubings attached to connec-
tion piece 9898 to the pump one after another, and pump
through ultrapure water.
Check whether solution emerges from connection piece 9898.
If one of the input or output leads remains blocked, replace
the connection piece (order number 6.2832.010).
4 Clean suppressor rotor
Clean the sealing surface of suppressor rotor 9999 using a lint-
free cloth and ethanol.
5 Insert suppressor rotor
Insert suppressor rotor 9999 in suppressor holder 100100 in such a
way that the tubing connections at the rear of the rotor fit in
the corresponding openings inside the rotor, and that one of
the three holes in the rotor can be seen from below in one of
the openings of the holder.
If the rotor has been inserted correctly, its sealing surface will
be about 4 mm inside the holder. If this is not the case, bring
the rotor into the correct position from below with the aid of a
sharp object (e.g. a screwdriver).
6 Clean connection piece
Clean the sealing surface of connection piece 9898 using a lint-
free cloth and ethanol.
100100999998989797
Fig. 24: Assembling the suppressor
9797 Screw nut 9999 Suppressor rotor (6.2832.000)
9898 Connection piece
(6.2832.010)
with input and output leads
100100 Suppressor holder