User guide

5.2 Maintenance and servicing
732 IC Detector / 733 IC Separation Center
141
2 Regenerate suppressor
Connect the inlet capillary 90 90 marked "H
2
SO
4
" to the 709 IC
Pump using the 6.2620.060 or 6.2744.040 coupling.
Set the flow at the 709 IC Pump to 0.5 mL/min and rinse the
suppressor unit with 1 mol/L H
2
SO
4
for 5 to 10 min.
As the pressure falls, slowly increase the flow at the 709 IC
Pump to 2 mL/min. Do not exceed a maximum pressure of
2 MPa (20 bar).
Switch suppressor to the next position using the <STEP>
key.
Connect the inlet capillary 9393 marked "H
2
O" to the 709 IC
Pump using coupling 6.2620.060 or 6.2744.040.
Set the flow at the 709 IC Pump to 0.5 mL/min and rinse the
suppressor unit with distilled H
2
O for 5 to 10 min.
As the pressure falls, slowly increase the flow at the 709 IC
Pump to 2 mL/min. Do not exceed a maximum pressure of
2 MPa (20 bar).
3 Connect suppressor to IC system
Reconnect the suppressor to the IC system and the 752
Pump Unit. If the pressure problems persist, replace the sup-
pressor rotor (see section 5.2.8).
5.2.7 Cleaning the suppressor
It may be necessary to clean the suppressor in the following cases:
High counterpressure on the suppressor connection
tubing
Irremediable blockage of the suppressor (the suppres-
sor can no longer deliver solutions)
Irremediable obstruction of the suppressor (the sup-
pressor can no longer be switched to next position)
To clean the connection piece and the rotor, proceed as follows (see
Fig. 24):
1 Disconnect suppressor from IC system
Disconnect input lead of suppressor from the separating
column and output lead to the detector.
2 Dismantle suppressor
Unscrew nut 9797 from suppressor holder 100100.
Pull out connection piece 9898 and suppressor rotor 9999 from
suppressor holder 100100 (the connection piece and the rotor
normally stick together).
Loosen connection piece 9898 from suppressor rotor 9999.