5.2 Maintenance and servicing
732 IC Detector / 733 IC Separation Center
139
4 Rinse with dist. H
2
O
Rinse IC system with dist. Or demin. Water (flow rate
1 mL/min) for ca. 10 min.
5 Rinse with eluent
Rinse IC system with the eluent needed for the separating
column you intend to install until the absolute conductivity
displayed on the 732 IC Detector is stable.
6 Remove coupling
Remove 6.2620.060 Coupling between column outlet capil-
lary 6767 and inlet capillary 8282 (see Fig. 16).
7 Connect new column to injector
Remove end caps from column 7676.
Screw inlet end of separating column 7676 (note flow direction)
to column connection capillary 6767 or to the precolumn (see
section 2.8.4/section 2.8.6).
8 Rinse column
Place beaker beneath the column outlet.
Switch on 709 IC Pump and rinse column with eluent for ca.
10 min, then switch off pump.
9 Connect column to detector block
Screw outlet end of separating column 7676 to inlet capillary 8282
or suppressor inlet capillary 8989.