User guide

5 Notes – Maintenance – Faults
732 IC Detector / 733 IC Separation Center
138
5.2.5 Changing separating columns
Identical separation system
If you wish to replace an IC separating column by a column of the same
type, proceed as follows (see Fig. 16 and Fig. 17):
1 Remove old column
Switch off pump drive of the 709 IC Pump.
Unscrew column from inlet capillary 8282 of the detector block
or from suppressor inlet capillary 8989.
Unscrew column from column connection capillary 6767 or the
precolumn.
2 Connect new column to injector
Remove end caps from column 7676.
Screw inlet end of separating column 7676 (note flow direction)
to column connection capillary 6767 or to the precolumn (see
section 2.8.4/section 2.8.6).
3 Rinse column
Place beaker beneath the column outlet.
Switch on 709 IC Pump and rinse column with eluent for ca.
10 min, then switch off pump.
4 Connect column to detector block
Screw outlet end of separating column 7676 to inlet capillary 8282
or suppressor inlet capillary 8989.
Changing the separation system
If you wish to replace an IC separating column by a column of a differ-
ent type, proceed as follows (see Fig. 16 and Fig. 17):
1 Remove old column
Switch off pump drive of 709 IC Pump.
Unscrew column from inlet capillary 8282 of the detector block
or from the suppressor inlet capillary 8989.
Unscrew column from column connection capillary 6767 or the
precolumn.
2 Connect detector block to injection valve
Connect column connection capillary 6767 using a 6.2620.060
Coupling directly to inlet capillary 8282 of the detector block
(see Fig. 16).
3 Rinse with HNO
3
(passivation)
Rinse IC system with c(HNO
3
) = 0.2 mol/L (flow rate
1 mL/min) for ca. 10 min.