User guide
4 Operation
732 IC Detector / 733 IC Separation Center
130
Method
METHOD
method name An.Dual
date: 1995-10-16 14:30:39
PARAM
>PARAM/detector
range: 100 µS/cm
full scale: 50.00 µS/cm
temp.coeff.: 2.5 %/°C
>PARAM/analog output
polarity: +
offset: 0 %fs
damping: off
>PARAM/Plot
auto start: off
time interval 1.0 s
time scale 10.0 mm/min
time scale label: rel
stop time off min
left: -5.000 µS/cm
right: 45.00 µS/cm
>PARAM/709 IC Pump
flow 0.80 mL/min
Pmax 7.0 MPa
Pmin 1.0 MPa
flow corr. 1.00
PROGRAM
progr.type: cycle
number of cycles 1
9.0 range: 100 µS/cm
full scale:5.00 µS/cm
mark: on
14.7 range: 100 µS/cm
full scale:50.0 µS/cm
mark: on
19.9 zero: on
49.9 valve A: fill
remote 0*******
pump R/S: off
50.0 flag: end
--------------
Switching of full-scale range to 5
µS/cm
Triggering of a marking signal
Switching of full-scale range to 50
µS/cm
Triggering of a marking signal
Triggering of the auto-zero function
Switching of valve A to "FILL"
Switching off of 752 Pump Unit
Switching off of 709 IC Pump
Program end
The program of the type "inject" is started each time injection valve A is
switched to the "INJECT" position. For the determination of nitrate and
phosphate, the full-scale range is briefly narrowed to 5 µS/cm and this
range marked by marking signals. After 20.0 min the IC system is ready
for the next injection. If there is no new program start within 50 min after
the last injection, valve A is automatically switched to the "FILL" position,
the 709 IC Pump and 752 Pump Unit are stopped and the program
ended.