User guide

4.7 Programming
732 IC Detector / 733 IC Separation Center
111
>PROGRAM/edit. XXX.X min
range: 100 µµS/cm
100, 200, 500 µS/cm
1, 2, 5, 10 mS/cm
Reset measuring range
This inquiry appears only with "FS change = on".
The measuring range is set to the inputted value (for
further details of the parameter "range", see sec-
tion 4.5.1).
If the measuring range just entered differs from the
measuring range valid at the start of the program, an
auto-zero function that is switched on will be switched
off. If the auto-zero function should be switched on
again, this can not be done in the same program step.
>PROGRAM/edit. XXX.X min
full scale: 100 µµS/cm
0.05 µS/cm10 mS/cm
Reset full-scale range
This inquiry appears only with "FS change = on".
The full-scale range is set to the inputted range (for
further details of the parameter "full scale", see
section 4.5.1).
>PROGRAM/edit. XXX.X min
zero: ---
---, on, off
Switch auto-zero on/off
This inquiry does not appear with "FS change = on".
on Trigger the auto-zero function (equivalent to
the <ZERO> key, see section 4.6.3).
off Switch off the auto-zero function (equivalent
to the <ZERO OFF> key, see section 4.6.4).
>PROGRAM/edit. XXX.X min
polarity: ---
---, +,
Reset the polarity of the output signal
+ Output of the positive full-scale range
(0+FS; see section 4.5.1).
Output of the negative full-scale range
(0FS; see section 4.5.1).
>PROGRAM/edit. XXX.X min
mark: ---
---, on
Trigger marking signal
on Output of a marking signal at the analog out-
put (equivalent to the <MARK> key, see
section 4.6.5).
>PROGRAM/edit. XXX.X min
valve A: ---
---, fill, inject
Switch injection valve A
This inquiry appears only with "SETUP/peripherals/operation
with 733 = on".
fill Switching of injection valve A to "FILL" posi-
tion (equivalent to the <FILL> key, see sec-
tion 4.6.1).
inject Switching of injection valve A to "INJECT"
position (equivalent to the <INJECT> key,
see section 4.6.2).