User guide
4 Operation
732 IC Detector / 733 IC Separation Center
108
<PROG R/S>
â
Program steps 1…k
â
return
â
Program steps m…n
â
end
No. of cycles
remote Single program with start via remote
A remote program is started externally via a
remote input line (see section 6.2.1) and
executed once only.
inject Single program with start via "INJECT"
An inject program is started each time injec-
tion valve A is switched to the "INJECT" posi-
tion and executed once only.
>PROGRAM/parameters
number of cycles 1
1…999
Number of cycles
This inquiry appears only with "PROGRAM/parameters/type =
cycle".
Number of executions of loop programs.
This parameter is identical to that of the same name in
the submenu ">CONFIG/auxiliaries" of the <CON-
FIG> key (see section 4.4.2). As access to the
<PROGRAM> key is blocked while a program is
running, the number of cycles can be changed only
there.
>PROGRAM/parameters
status: disabled
enabled, disabled, test
Program status
The three following program statuses are possible:
enabled Program enabled
Irrespective of the program type, a pro-
gram can be started only when it is en-
abled. This status is shown by the lighting
up of the green LED 55 in the <PROG
R/S> key. If a method with program
steps is loaded, the program status is
automatically set to "enabled".
disabled Program disabled
A disabled program can not be started. In
this status, the green LED 55 in the
<PROG R/S> key remains off. When the
instrument is switched on, the program
status is always set to "disabled".