User guide
4.4 Basic settings
732 IC Detector / 733 IC Separation Center
95
>CONFIG
>CONFIG/733 IC Sep.Cent.
Settings for 733 IC Separation Center
This submenu appears only with "operation with 733 = on".
>CONFIG/733 IC Sep.Cent.
valve A
display only
Configuration 733 IC Separation Center
Depending on the instrument configuration, the follow-
ing messages appear:
valve A 733.0010 (1 injector).
valve A + valve B 733.0020 (2 injectors).
valve A + suppressor 733.0030 (1 injector + 1
suppressor module).
733 not accessible 733 not connected or
cable connection faulty.
>CONFIG/733 IC Sep.Cent.
control: no restriction
no restriction, 732 only
Valve control of the 733 IC Separation
Center
no restriction Suppressor module and valves
can be operated with the <FILL>
and <INJECT> or <STEP> keys
at 733 IC Separation Center and
also via the 732 IC Detector.
732 only Suppressor module and valves
can be operated only via 732 IC
Detector (the keys on the 733 are
locked).
>CONFIG/733 IC Sep.Cent.
>CONFIG/733/valve A
submenu
Settings for injection valve A
>CONFIG/733/valve A
status fill
display only
Display of the valve setting
Depending on the position of the valve, one of the
following messages appears:
fill Filling of sample loop.
inject Injection of loop contents.
undefined Undefined position.
>CONFIG/733/valve A
trigger: fill
inject, fill
Switching of the valve
After confirmation with <ENTER>, the is moved
to the selected position:
fill Filling of sample loop.
inject Injection of loop contents.