User Manual

3.4 BuretBuret units
726 Titroprocessor, Instructions for use
5050
3.4.3 685 Dosimat, exchange units
For the 685 Dosimat exchange units the types of buret units are not defined
as tubing dimensions. In this case the rinsing volume and the number of
rinsing cycles are entered .
The same names can be used as for the dosing units of the 700 Dosino.
The 726 Titroprocessor can determine the type of dosing unit itself.
ÚÄÄÄConfiguration ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ(Program version 5.726.0012)ÄÄÄÄÄÄÄÄ¿
³ ³
³ DialÚÄÄÄ Tubings ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ¿ ³
³ ³ ³ ³
³ LCD ³ Dosimat 685: Exchange units, preparation ³ ³
³ LCD ³ ³ ³
³ ³ # Type Volume Cycles Dosing rate Filling rate ³ ³
³ Fehl³ ÄÄ ÄÄÄÄÄÄÄÄ ÄÄÄÄÄÄÄÄÄÄ ÄÄÄÄÄÄ ÄÄÄÄÄÄÄÄÄÄÄÄÄ ÄÄÄÄÄÄÄÄÄÄÄÄÄ ³ ³
³ Edit³ 1 Default 3.5 mL 2 max mL/min max mL/min ³ ³
³ Edit³ 2 Changer 10.0 mL 2 max mL/min max mL/min ³ ³
³ ³ 3 mL mL/min mL/min ³ ³
³ ³ 4 mL mL/min mL/min ³ ³
³ ³ 5 mL mL/min mL/min ³ ³
³ ³ 6 mL mL/min mL/min ³ ³
³ ³ 7 mL mL/min mL/min ³ ³
³ ³ 8 mL mL/min mL/min ³ ³
³ ³ 9 mL mL/min mL/min ³ ³
³ ³ 10 mL mL/min mL/min ³ ³
³ ³ ³ ³
³ ³ ³ ³
³ ³ ³ ³
³ ÀÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ QUIT ÄÙ ³
³ ³
ÀÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ QUIT ÄÙ
ÚÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄ¿ ÚÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄ¿ ÚÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄ¿
³ ³ ³ ³ ³ ³ ³ 700 ³ ³Delete ³Delete ³ Check ³
³ ³ ³ ³ ³ ³ ³Dosino ³ ³line ³ all ³ type ³
ÀÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÙ ÀÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÙ ÀÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÙ
Type Name of the buret unit. Two types are already predefined.
(Default=standard settings, see above,
Changer=standard setting for sample changer
applications, see above.)
Volume Reagent volume for one rinsing cycle.
Cycles Number of rinsing cycles to be carried out.
Dosing
rate The maximum value for the dosing speed depends on the
cylinder size of the buret used.
Buret cylinder Max. speed
1 mL 3 mL/min
2 mL 6 mL/min
5 mL 15 mL/min
10 mL 30 mL/min
20 mL 60 mL/min
50 mL 150 mL/min
For smaller buret cylinders a value of up to 150 mL/min
can also be entered, but when the function is carried out
the dosing speed will automatically be reduced to the