User Manual

3. Configuration
726 Titroprocessor, Instructions for use
3333
3 Configuration
In order to guarantee problem-free work with the 726 Titroprocessor it is
essential that the basic settings of all instrument components are
configured according to your requirements.
3.1 Basic settings
By pressing the softkey [Config.] the basic configuration can be
accessed from the main page.
ÚÄÄÄConfiguration ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ(Program version 5.726.0012)ÄÄÄÄÄÄÄÄ¿
³ ³
³ Dialog in english Date 1997-12-04 ³
³ Time 16:05:52 ³
³ LCD contrast 100 ³
³ Screen save 30 min Standard data file name DATA ³
³ File name type <Ident1>.### ³
³ Error beep on File comment type Ident1,2+Meth+Remark ³
³ Edit beep on ³
³ Edit mode insert Power on PREP warning on ³
³ ³
³ Remote terminal mode off ³
³ File transfer protocol Metrohm(Bkup) ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
ÀÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ QUIT ÄÙ
ÚÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄ¿ ÚÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄ¿ ÚÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄ¿
³Device ³File ³Access ³ ³Sample ³Inter- ³Config.³ ³Tubings³ ³ >> ³
³header ³ ³control³ ³changer³faces ³idents.³ ³ ³ ³ >> ³
ÀÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÙ ÀÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÙ ÀÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÙ
The basic settings for the Titroprocessor 726 can be configured on this
dialog page. The softkeys allow access to the configuration of further
components.
Dialog in Dialog language. English, German, French and
Spanish are available. After changing the dialog
language the instrument should be switched off and
then back on again.
LCD contrast Contrastcan be selected in 250 steps.
The contrast can also be set with the key
combinations <ALT><éé> and <ALT><êê>.
Screen save Delay period for switching the screen off. If no key
on the Titroprocessor is pressed within the set period
then the background illumination of the LCD screen
will be automatically switched off (for its own care).
Pressing a key switches it on again.
Error beep Acoustic error message signal on/off
Edit beep Acoustic editing error message signal on/off
Edit mode Switches between overwrite/insert mode during data
input