User Manual

9.1 Results page
726 Titroprocessor, Instructions for use
212212
9 Results and statistics
When a determination has been completed the calculations which have
been defined in the method are carried out. The results are shown on the
results page. The evaluation of the measuring data can be carried out at a
later data with altered evaluation criteria. Sample data and calculation
formulas can also be altered for later calculations.
Alterations which are carried out for a later evaluation or
calculation are not automatically incorporated in the method.
Subsequent determinations will still be carried out with the
settings defined in the original method!
The corresponding method does not need to be loaded in order for
later re-evaluations to be carried out.
9.1 Results page
The results page is called up from the main page with the [Results]
softkey.
ÚÄÄÄ Results: HALIDES.001 Original data ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ¿
³ # 3 DET_U* Determination of halides ³
³ EP1 -359.7 mV 1.029 mL vol.iodide 1.029 mL ³
³ EP2 -146.9 mV 4.700 mL vol.bromide 3.671 mL ³
³ EP3 169.6 mV 7.510 mL vol.chloride 2.809 mL ³
³ iodide 81.78 mg/L ³
³ bromide 183.72 mg/L ³
³ chloride 62.38 mg/L ³
³ ³
³ Error list: ³
³ ÄÄÄÄÄÄÄÄÄÄÄ ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
ÀÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ QUIT ÄÙ
ÚÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄ¿ ÚÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄ¿ ÚÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄ¿
³Sample ³Select ³Determ.³ ³ * ³ * ³ * ³ ³Local ³Calib. ³ >> ³
³data ³ ³report ³ ³ Param ³Formula³ Report³ ³ComVars³data ³ >> ³
ÀÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÙ ÀÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÙ ÀÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÙ
The second softkey bar:
ÚÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄ¿ ÚÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄ¿ ÚÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄ¿
³Extend.³ File ³Statisij ³ ³Recalc ³Recalc ³ ³ ³ ³ >> ³
³values ³ ³tics ³ ³ ³ all ³ one ³ ³ ³ ³ >> ³
ÀÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÙ ÀÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÙ ÀÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÙ
Bar cursor
Name of the
determination file
Result attribute
Titration or measuring
mode with command
line and comments
EP list with measured value and EP
volume
Results list with comments
List of errors
which have
occurred