Manual

4.4 XDOS mode - controlled dosing
711 Liquino, Instructions for use
82
Curve Report
711 Liquino 0220 711.0011
date 1998-09-30 time 11:39
*** Dev. Dept.
*** dm
*** Metrohm Ltd
mode: XDOS user method: TempDos1
lot no. 12-2
============
volume/temperature curve
Combined List
711 Liquino 0220 711.0011
date 1998-09-30 time 11:39
*** Dev. Dept.
*** dm
*** Metrohm Ltd
mode: XDOS user method: TempDos1
lot no. 12-2
time vol.(mL) temp(°C) event
----------------------------------------
000:00:00 0.000 *** START ***
000:00:00 0.000 27.5
000:00:10 0.828 28.5
000:00:20 1.656 29.5
000:00:25 2.075 T-limit,wait
000:00:30 2.075 30.0 upper T-limit
000:00:40 2.075 30.0 upper T-limit
000:00:42 2.075 continue
000:00:50 2.751 29.5
000:01:00 3.584 29.5
000:01:10 4.416 29.5
000:01:20 5.249 29.0
000:01:30 6.081 29.5
000:01:40 6.914 29.5
000:01:50 7.746 29.0
000:02:00 8.579 29.0
000:02:10 9.411 28.5
000:02:17 10.000 *** END ***
000:02:20 28.5
000:02:30 28.5
============
tabular list of volumes, temperatures and
events
start of dosing
temperature limit reached, dosing interrupted
temperature back in limit, dosing continued
end of dosing
procedure ended after dosing unit has been filled.
Temperature curve
Volume curve
Temperature limit
Temperature
limit reached
Dosing interrupted