Manual

4. Detailed description
711 Liquino, Instructions for use
81
4.4.2 Examples of reports - XDOS mode
Temperature-controlled dosing with volume and time selection
Parameter Report
711 Liquino 0220 711.0011
date 1998-09-30 time 11:51
*** Dev. Dept.
*** dm
*** Metrohm Ltd
mode: XDOS user method: TempDos1
lot no. 12-2
parameters
>XDOS parameter
XDOS type: volume&time
volume 10.0 mL
dosing time 00:02:00
temp. measuring: on
rec. interval 00:00:10
fill rate max mL/min
auto fill: on
>XDOS dosing units
setup: single
dosing drive: DOS1
dos.unit: D3
port use: custom
dosing port 1
filling port 2
waste port 1
>XDOS monitoring
temperature: on
low limit 20.0 °C
upper limit 30.0 °C
action: wait
alarm pins: none
volume: off
>XDOS preselections
req.ident: id1
>>XDOS prompts
id1 lot no.
id2 id2
id3 id3
start delay 00:00:00
>XDOS reports
temp. auto scale: on
report1: result
report2: curve
report3: comb. list
report4: off
report5: off
report6: off
report7: off
report8: off
------------
instrument number and program version
date and time
3-line report header
mode and method name
ID1, identification with preselected display text
dosing parameters
dosing unit settings
monitoring function settings
request settings
report definitions
identification of reproduced reports
Result Report
711 Liquino 0220 711.0011
date 1998-09-30 time 11:39
*** Dev. Dept.
*** dm
*** Metrohm Ltd
mode: XDOS user method: TempDos1
lot no. 12-2
total dosing time = 000:02:17
total volume = 10.000 mL
dosing rate = auto
============
see above
effective time *
dosed volume
dosing rate
identification of original report
* total time including waiting times during infringements of temperature limits or during filling