Manual

4. Detailed description
Liquino 711, Instructions for use
71
>>er r . l i mi t s 20mL- cyl .
Submenu for 20 mL dosing units
er r or max. vol . +/ - 60 µL
1…999 µL maximum volume error
sl ope +/ - 0. 003
0.001…0.1 maximum variation in slope
i nt er cept +/ - 30 µL
1…99 µL maximum y-axis intercept
>>er r . l i mi t s 50mL- cyl .
Submenu for 50 mL dosing units
er r or max. vol . +/ - 150 µL
1…999 µL maximum volume error
sl ope +/ - 0. 003
0.001…0.1 maximum variation in slope
i nt er cept +/ - 50 µL
1…99 µL maximum y-axis intercept
4.3.3 Keyboard and graphics options
Further configuration possibilities are available:
Locking the keyboard when the Liquino 711 is operated exclusively by
remote control.
Locking individual dialog ranges
Settings for graphics printout
Hiding individual instrument modes
These settings can be made in a special setup menu which is only accessi-
ble when the Liquino is switched on with the <CONFIG> key pressed
down.
CONFIG
A
Switch the Liquino off, press down the <CONFIG> key and keep this
pressed down while you switch the instrument on again.
Main menu:
set up
>keyboar d opt i ons
open submenu with <ENTER>
set up
>gr aphi cs pr i nt out
use <> or <> to move up or down by 1 menu item
set up
>show/ hi de modes
use <HOME> or <END> to jump to first or last menu item
use <QUIT> to return to the basic condition
Submenus:
set up
>keyboar d opt i ons
Locks keyboard or particular keys
Navigate with <ENTER>
>keyboar d opt i ons
l ock keyboar d: of f
Switches off keyboard
on
off
= the Liquino cam only be operated via external software.
If the keyboard is locked the setup menu is still available to switch off the
keyboard lock.