4. Detailed description
Liquino 711, Instructions for use
69
4.3.2 The configuration menu
Please note: menu entries with a colon offer a selection list. Press the <SELECT> key to make your se-
lection.
A
CONFIG
>peripheral units
>auxiliaries
>manual
>dosing units
>RS232 settings
>report header
>GLP error limits
Di spl ay I ni t i al val ue
Input range Explanation
>per i pher al uni t s
Type of instrument connected (RS232 interface)
The specification of the externally connected instruments is extremely important for the correct functioning of the Liquino in an
automated system. Instead of a printer a computer or other instrument (e.g. a Metrohm titrator) can be connected to the serial
RS232 interface, in order to store Liquino reports or realize complex remote control functions. The is done by selecting the setting
'
send t o: I BM'.
If your type of printer is not listed select 'IBM' or 'Epson'; for laser printers select 'HP'. If your balance is not listed select 'Precisa'.
send t o: I BM
IBM, Epson, Seiko, Citizen, HP printer type
bal ance: Sa r t or i us
Sartorius, Mettler, Mettler AT,
AND, Precisa
balance manufacturer
>auxi l i ar i es
Basic Liquino settings
The configuration settings are retained even when the instrument is switched off.
di al og: eng
lish
english, german dialog language
dat e 1998- 09- 11
JJJJ-MM-TT system date
t i me 08: 10: 15
SS:mm:ss system time
devi ce l abel XXXXXXXX
8 ASCII characters instrument identification
pr ogr am 711. 0011
not editable software version number
di spl ay cont r ast 3
0…7 LCD display contrast
beeper : on
on; off beeper for warnings and error messages
PREP r emi nder : on
on; off automatic request to prepare the dosing units.
>manual
Settings for manual operation
These settings are only effective in manual operation, i.e. for dosing and dispensing.
In order for the PREP and EMPTY functions to work properly the type of dosing device must be given for each dosing device which
is connected. When the type of dosing unit is specified the corresponding tubing dimensions as defined under 'Configura-
tion>Dosing units' will be used for the internal calculation of the rinsing volume.
By navigating with the <ENTER> key the settings for all 4 dosing device connections can be made one after the other.
dos. uni t : def aul t
selection list, see '>dosi ng
uni t s
'
type of dosing unit connected.
fill rate ma x mL / mi n
0.001…166.666 mL/min; max the maximum filling rate depends on the volume of the
dosing cylinder.
aut o f i l l : of f
on; off automatic filling after dosing.
>>dosi ng
Submenu for dosing behavior
r at e t ype: i ncr easi ng
increasing; constant increases dosing rate for precise manual dosing .
sl ope 4
(only for increasing dosing rate)
1…10 slope of increase in dosing rate.