3.2.2 Defining dosing units
In order for the complete tubing system of a Dosino to be fully filled with the
reagent solution without any air bubbles the 711 Liquino must know the di-
mensions of the connected tubing. For this reason a dosing unit can be de-
fined in the configuration menu and then selected for the method.
*)
Except for the "Sampling dosing unit" in the SMPL mode the standard port
occupancy for all modes is as follows:
Port 1: dosing unit output
Port 2 (bottle): refilling opening
CONFIG
A
• Press <CONFIG> and then 3x <
>.
Di spl ay: conf i gur at i on
>dosi ng uni t s
ENTER
YES
•
Access the submenu ‘dosing units’ with <ENTER>.
Di spl ay: > dosi ng uni t s
dos. uni t : D1
DOS NR
SELECT
P
• One of the ten configurable dosing units can be selected with the
<SELECT> key. Press <ENTER> to access the settings for the corre-
sponding dosing unit. The index for the dosing unit is located in the top
right-hand corner of the display.
Di spl ay: >dosi ng uni t s 1
I D D1
• Enter an identification text for the dosing unit or leave the preset entry
as it is. Text input is explained on page 63.
• Accept the entry with <ENTER>.
Di spl ay: >dosi ng uni t s 1
l engt h t ube1 400 mm
• Now enter the length of tubing at Dosino port 1 (for tandem dosing
including the tubing length up to the T-piece) and accept the entry with
<ENTER>.
Di spl ay: >dosi ng uni t s 1
di am. t ube1 2. 0 mm
• The inner diameter of the tubing at Dosino port 1 can be entered here
and confirmed with <ENTER>.
Di spl ay: >dosi ng uni t s 1
l engt h t ube2 250 mm
• Now enter the length of the tubing at Dosino port 2 (bottle) and accept
the entry with <ENTER>.
Di spl ay: >dosi ng uni t s 1
*)
In a single method ‘standard’ can also be selected for the dosing unit used. For the modes XDOS, CONT
and GLP ‘standard’ means:
Port1 (dosing port): L = 40cm, D = 2mm
Port2 (bottle): L = 25cm, D = 2mm.
For PIP ‘standard’ means: Port1 (pipetting tubing): L = 150cm, D = 2mm
Port2 (bottle): L = 25cm, D = 2mm.
For SMPL ‘standard’ means: Port1 (sample tubing): L = 40cm, D = 2mm
Port2 (measuring cell): L = 16cm, D = 2mm