2.5 Connections
711 Liquino, Instructions for use
14
Example of a sample changer method:
>st ar t sequence
1 CTL: Rm: INIT
>sampl e sequence
1 CTL: Rm: START devi ce1
2 SCN: Rm: r eady1
3 Lift: 1: shift mm
4 CTL:Rm: *********1****
5 Wai t 1 s
6 CTL:Rm: *********0****
7 Move: 1: sampl e
8 Li f t : 1: wor k mm
9 CTL:Rm: *********1****
10 Wai t 1 s
11 CTL:Rm: *********0****
12 SCN: Rm: r eady1
13 Move: 1: spec. 1
14 Li f t : 1: wor k mm
15 CTL:Rm: *********1****
16 Wai t 1 s
17 CTL:Rm: *********0****
18 SCN: Rm: end1
19 St i r : 1: ON s
20 CTL: Rm: START devi ce2
2 1 SCN: Rm: e nd2
22 St i r : 1: OFF s
23 Pump: 1. 1: 12 s
24 Pump: 1. * : 10 s
25 Pump: 1. 1: 12 s
28 NOP
>f i nal sequence
1 Lift: 1: shift mm
2 NOP
Initialize remote interface
Start Liquino (PIP mode)
Wait for pipetting step
Lift out of sample vessel
Switch pipetting further (corresponds to
<ENTER>)
separating bubble is aspirated
Reset remote interface
Position sample beaker
Immerse pipet tubing in sample
Switch pipetting further (corresponds to
<ENTER>)
sample is aspirated
Reset remote interface
Wait for next pipetting step
Move to titrating beaker
Pipet tubing in titrating beaker
Switch pipetting further (corresponds to
<ENTER>)
sample is ejected
Reset remote interface
Wait for end of PIP mode
Stirrer on
Start Titrino
Wait for end of determination
Switch off stirrer
Aspirate titration beaker
Rinse titration beaker
Aspirate titration beaker
711 Liquino731 Relay Box 772 Pump Unit Heating Bath
Liquino with 772 Pump Unit as delivery or aspiration pump and a heating
bath.
711
731
772
With the aid of a Metrohm 731 Relay Box the output leads of the 711 Liquino
can be monitored and, according to their line conditions, two low-voltage
d.c. voltage outputs and two mains voltage outputs can be switched.
In this way the heating bath can be switched automatically via the tempera-
ture monitoring function. If the measured temperature falls below a certain
value the bath is heated up.
A connected pump (e.g. the Metrohm 772 Pump Unit) can be selectively
switched on or off during the method procedure.
cable
6.2125.100
Dosinos
heating bath
mains cable