Manual

2.5 Connections
711 Liquino, Instructions for use
12
711
Liquino Titrino 683 Pump 730 Sample Changer
Cable 6.2141.100 is used for the SMPL mode if aspiration and rinsing the ti-
tration vessel are to be carried out as quickly as possible (with 2 x Pump
683). The sample changer connection can be left open if no sample chang-
er is required.
The Titrino, the pumps and the sample changer can also be addressed with
the control commands of a SEQ method, for example:
Start Titrino
>SEQ sequence st ep 4 ( CONT ROL command)
pat t er n * 1* * ---- starts Titrino (starts pulse)
>SEQ sequence st ep 5 ( Pause command)
pause 00: 00: 01
wait 1 s
>SEQ sequence st ep 6 ( CONT ROL command)
pat t er n * 0* * ----
starts Titrino (ends pulse)
>SEQ sequence st ep 7 ( SCAN command)
pattern ****1---
wait for ready message from Titrino
Run pump 1 for 10 s
>SEQ sequence st ep 8 ( CONT ROL command)
pattern ***1---- switches on 1st 683 pump
>SEQ sequence st ep 9 ( Pause command)
pause 00: 00: 10
wait 10 s
>SEQ sequence st ep 10 ( CONT ROL command)
pattern ***0----
switches off 1st pump 683 again
Run pump 2 for 1 min 5 s
>SEQ sequence st ep 11 ( CONT ROL command)
pat t er n * * 1* ---- switches on 2nd pump 683
>SEQ sequence st ep 12 ( Pause command)
pause 00: 01: 05
wait 1 min 5 s
>SEQ sequence st ep 13 ( CONT ROL command)
pat t er n * * 0* ----
switches off 2nd pump 683 again
B
A
Titrino
C
D
230
683 Pump Unit
I
0
683 Pump Unit
I
0
730
cable 6.2141.100