6.2 Error messages
711 Liquino, Instructions for use
146
Error messages relating to data transmission:
Note:
If neither a printer nor a computer is connected the report output
must be switched off.
General form of error message:
*** RS error xx Error during data transfer or printing. Check whether the connection
cable has been attached correctly and that the RS232 parameters of
the transmitting and receiving instruments are the same. See
<CONFIG>, >RS232 settings
Reception errors:
*** RS error 36 Parity error
Exit: <QUIT> and set the same parameter on both instruments.
*** RS error 37 Framing error. The receiving instrument may not be ready to receive
data. Exit: <QUIT> or switch Liquino off and on. Check settings of
receiving instrument.
*** RS error 38 Overrun error. At least 1 character could not be read.
Exit: <QUIT>.
*** RS error 39 The internal processing buffer in the Liquino is overloaded (>128
characters). Exit: <QUIT>.
Transmission errors:
*** RS error 42 CTS=OFF. Handshake not satisfied for more than 1 s.
Exit: <QUIT>. Is the receiver switched on and ready to receive?
*** RS error 43 The Liquino transmission has been interrupted by XOFF for at least 6
s. Exit: <QUIT>.
*** RS error 45 The receive buffer of the Liquino contains an incomplete character
string (L
F
is missing). The Liquino transmission is therefore blocked.
Exit: send L
F
or <QUIT>.
Warnings
* * * Dosi ng not cont i nuous. Lower r at e * * *
You have defined a dosing rate which is too high for tandem dosing. Inter-
ruption-free dosing is not possible as a filling process (including stopcock
switching) takes longer than dosing a cylinder volume. Press <HOLD> and
reduce the dosing rate to 85% at the most of the maximum possible dosing
rate for the dosing unit. Dosing can then be continued with
<HOLD/CONTINUE>.
*** Rate to hi gh. Adj ust par amet er ! * * *
The necessary dosing rate required to fulfill the task you have defined can-
not be reached. Press <HOLD> and reduce the volume to be dosed in or
extend the dosing period. Dosing can then be continued with
<HOLD/CONTINUE>. If you do not correct the parameters dosing will be
carried out at the highest possible dosing rate. However, the defined dosing
time cannot be observed in this case.