Manual

5.2 Please note …
711 Liquino, Instructions for use
136
chloromethane, trinitrophenol-2,3,6, vinyl acetate, vinyl chloride, vinyli-
dene chloride
Never
dose hot (>60°C) chemicals or solutions with Dosinos!
Metrohm offers tubing made of FEP (tetrafluoroethylene-
hexafluoropropylene copolymer) and PTFE (polytetrafluoroethylene).
These materials are largely resistant to organic solvents, strong acids
and alkalis.
FEP tubing is not suitable for:
perchloric acid
fluorine
FEP tubing has a limited resistance to:
chlorosulfonic acid
FEP tubing has a limited resistance at higher temperatures to:
crude oil products
perchloroethylene
tetrachloroethylene
PTFE tubing has a limited resistance to:
halogenated hydrocarbons
perchloric acid
fluorine
PTFE tubing has a limited resistance at higher temperatures to:
crude oil products
perchloroethylene
tetrachloroethylene
Unoccupied dosing unit connections (Ports) can be sealed with
drying tubes or stoppers.
Divide your synthesis run into part steps and create separate XDOS
methods for the individual dosing steps. Then you can link together the
individual part steps (methods) to form a sequence (SEQ mode) and
add further commands to it.