Manual

4. Detailed description
711 Liquino, Instructions for use
125
4.9.2 Example of report - GLP mode
Validation of dosing units
Parameter Report
711 Liquino 0220 711.0011
date 1998-10-03 time 11:32 - 11:35
*** Dev. Dept.
*** dm
*** Metrohm Ltd
mode: GLP user method: GLP test
operator D. Miller
drive id Dosino1
dos. unit id Dos20_1
parameters
>GLP parameter
dosing drive: DOS2
dos.unit: default
temperature 22.5 °C
>GLP options
liquid: water
weight from: balance
balance id
>GLP reports
report1: result
report2: parameter
report3: curve
report4: detail
report5: off
report6: off
report7: off
>>err. limits 20mL-cyl.
error max.vol.+/- 60 µL
slope +/- 0.003
intercept +/- 30 µL
============
instrument number and program version
date and time
3-line report header
mode and method name
request texts
parameters
options
report definitions
configured error limits
identification for reproduced reports
Result Report
711 Liquino 0220 711.0011
date 1998-10-03 time 11:32 - 11:35
*** Dev. Dept.
*** dm
*** Metrohm Ltd
mode: GLP user method: GLP test
operator D. Miller
drive id Dosino1
dos. unit id Dos20_1
dosing unit size 20 mL
temperature 22.50 °C
liquid water
>evaluation
intercept (<30) -11.2 uL ok
slope (+/-0.003) 1.00076 ok
error max. vol. (<60)-2 uL ok
-> GLP test PASSED!
============
instrument number and program version
date and time
3-line report header
mode and method name
request texts
parameters
results (with limits)
result