4. Detailed description
711 Liquino, Instructions for use
111
Rinse Dosino
rinse
Dosino
repeat
rinsing
empty
Dosino
The Dosino aspirates the sample solution. The aspi-
rated amount is defined by the sample rinsing volume
and the tubing volume at port 1.
The dosing cylinder is emptied into the waste bottle
via port 4.
This initial rinsing can be repeated up to 10 times.
Add auxiliary solution
add aux.
solutions
With further Dosinos up to three different auxiliary
solutions can be added, see '>SMPL Dosing units'.
After the addition of each auxiliary solution a waiting
period can be programmed (in HH:MM:SS). The ad-
dition of auxiliary solutions is optional. It is possible
before or after sample transfer.
Transfer sample
aspirate
sample
add
sample
The Dosino aspirates the sample via port 1.
The defined sample volume is dosed into the measur-
ing cell (via port 2).
Dilute sample
aspirate
solvent
add
solvent
The Dosino aspirates the solvent via port 3.
The defined dilution volume is dosed into the measur-
ing cell (port 2).