Manual

4. Detailed description
711 Liquino, Instructions for use
105
4.7.2 Example of a report - PIP mode
Pipetting with dilution
Parameter Report
711 Liquino 0220 711.0011
date 1998-10-02 time 13:22
*** Dev. Dept.
*** dm
*** Metrohm Ltd
mode: PIP user method: Pip05
parameters
>PIP parameters
pipetting vol. 0.5 mL
diluting vol. 9.5 mL
air gap 10 mm
>PIP dosing unit
dosing drive: DOS1
dos.unit: default
>PIP rates
feed rate max mL/min
fill rate max mL/min
>PIP preselections
req. pip. vol.: on
>PIP reports
report1: result
report2: off
report3: off
------------
instrument number and program version
date and time
3-line report header
mode and method name
pipetting parameters
dosing unit settings
dosing and filling rate settings
request settings
report definitions
identification for reproduced report
Result Report
1998-10-02 13:24 cylinder 10mL
pip. vol. 0.500mL dil. vol. 9.500mL