4.7 PIP mode - precise pipetting and diluting
711 Liquino, Instructions for use
104
par amet er s
>PI P r at es
Dosing and filling rate settings
Navigate with <ENTER>
>PI P r at es
f eed r at e ma x mL / mi n
Dosing rate
(should not exceed 20 mL/min.)
0.001…166.66,
max
= highest possible dosing rate (depends on type of
dosing unit.)
>PI P r at es
fill rate ma x mL / mi n
Filling rate of dosing unit
(should not exceed 20 mL/min.)
0.001…166.66,
max
= highest possible filling rate (depends on type of dosing
unit.)
par amet er s
>PI P pr esel ect i ons
Definition of requests at start
Navigate with <ENTER>
>PI P pr esel ect i ons
req. pip. vol.: of f
Automatic request for pipetting volume
on, off
par amet er s
>PI P r e por t s
Report definitions for result output
Navigate with <ENTER>
>PI P r epor t s no. 1
r epor t : of f
Selection of one or more reports to be printed
out when dosing process is completed.
off
result,
parameter,
ff
= no report
= list of relevant results
= list of all method parameters
= FORM FEED; triggers page feed at printer
(important for laser and inkjet printers, for
ejecting pages)
>PI P r epor t s no. 2
r epor t : of f
… up to three reports are possible
An example of a report is shown on the following page.
Maximum values for
dosing and filling rates
Cylinder vol. Max. rate
2 mL 6.66 mL/min
5 mL 16.66 mL/min
10 mL 33.33 mL/min
20 mL 66.66 mL/min
These are pure piston
stroke rates, without taking
the filling times into ac-
count.