User manual

35Wattmeters
6. MAINTENANCE
Only use specified spare parts for maintenance purposes. The manufacturer cannot
accept any responsibility for accidents occurring following repairs carried out outside
its after-sales department or approved maintenance network.
6.1 Replacing the batteries
Disconnect the leads from the instrument
Press the OFF key
Unscrew the “quarter-turn screw” located at the top of the battery and fuse access hatch
(on the back of the instrument) and remove the hatch by pulling the stand.
Replace the 6 batteries located on either side of the opening with 6LR6-type 1.5 V alkaline
batteries, taking care to comply with the polarities indicated on the printed circuit.
Reinstall the hatch and lock the quarter-turn screw.
6.2 Replacing the fuse
Disconnect the leads from the instrument
Press the OFF key
Proceed as indicated in § 6.1 for replacement of the batteries.
Replace the fuse with a 10 A fuse of the same type (10 A - 200 kA - 10 x 38 - 600 V -
CCMR class).
Reinstall the hatch and lock the quarter-turn screw.
6.2 Storage
If the wattmeter is not used for a period of more than 60 days, remove the batteries and store
them separately.
6.3 Cleaning
Disconnect the leads from the instrument and press the OFF key
Clean the unit with a cloth slightly moistened with soapy water. Clean off with a damp
cloth. Then dry quickly with a cloth or pulsed air.
Chapter VI