User manual

29Wattmeters
L1
W
L1
L2
L2
L3 L3
U U
I
I
Fig 4.4.2 Three-phase/balanced wires (T3FE)
connection diagram (PX 120 only)
3.4.1Specific reference conditions
DC: AC component < 0.1% of the DC signal
AC: - Sinusoidal signals
- PF = 1
- Frequency: 50 Hz
3.4.2 Active power measurements
1. Press the ON/OFF key: the first screen appears
2. Connect the measurement leads to the instrument’s current measurement terminals
(on the left), complying with the polarities indicated: red lead on the +terminal and
black lead on the “COM” terminal.
Connect the measurement leads to the instrument’s voltage measurement terminals
(on the right), complying with the polarities indicated: red lead on the +terminal
and black lead on the COMterminal.
Set up as indicated in the connection diagrams in Fig.4.4.1 (single phase) or Fig.4.4.2
(balanced three-phase, T3FE only PX 120), depending on the case, making sure if
possible that the maximum acceptable limits are not exceeded (see table below).
Range switching is automatic.
Characteristics
Display range 1000 W 6 kW
Measuring range 10.00 W...999.9 W 1000 W...5999 kW
Accuracy 1% of reading ±2 cts 1% of reading ±2 cts
Accuracy for DC 2% of reading ±3 cts 2% of reading ±3 cts
Resolution 0.1 W 1 W
Display stability 5 counts in normal mode
2 counts in SMOOTH mode
Response time 400 ms in normal mode
of the display 3 s in single-phase, in SMOOTH mode
Note 1
:
The measurement of the active power is a signed measurement.
Above 6 kW +10%, the display indicates OL”.
Note 2: In three-phase, the active power is only measured on the fundamental
components so it is only valid for sinusoidal signals.
Chapter III