6.3.2. Electromagnetic compatibility and susceptibility
Compliant with IEC 61326-1 for emissions and immunity in industrial use.
6.3.3. Protection from electric shocks
Type A clamp with double insulation or reinforced insulation in the grip part in
normal use, and double insulation or reinforced insulation between the primary
and the secondary output and the terminations part of the USB power supply and
the BNC output.
Maximum service voltage per EN61010-2-032
- 300V in installations of category III and degree of pollution 2.
- 600V in installations of category II and degree of pollution 2.
6.3.4. Variations in the nominal of the use
Quantity of influence
Influence
Typical
Maximum
USB supply voltage
5 VDC ±5%
or storage battery voltage
between 20% and
100% charge
0
none
Relative humidity
10% at 90% RH
0.1% of Is
Temperature on
measurement
-10°C to +45°C
±0.12%/°C
45°C to +50°C
±0.22%/°C
Settling time
upon start of operation
10 s
Pass band (without
filter)
DC to 1kHz
1%
to 100kHz
-0.2dB
-0.5dB
to 500kHz
-0.5dB
-1dB
to 1MHz
-1.5dB
-3dB
Position of the cable
DC to 100Hz
±1.5%
Repeatability of the
measurement
10 open./clos.
mean : 0.3%
Common mode
rejection
50Hz
>90dB or 0.003%
400Hz
>80dB or 0.01%
Rejection of an external
current at 50Hz (
1
)
near air gap
30dB
jaws
34dB
grip
40dB
External magnetic field
(spatial position) (
2
)
DC (Earth's)
±0.3mV or
±30mA
±1mV or
±100mA
High-frequency
electromagnetic field,
10V/m
between 650MHz and
750MHz
value of the current
1.2A
1.5A max
(
1
) Conductor outside the jaws and parallel to the position of a conductor in the jaws.
(
2
) Measurement made after demagnetization of the clamp. For this purpose, set to AC, I = 0, gradually
increase the value to I = 40A. Gradually reduce the value to I = 0.