Instructions
Fernzugriff 
95 
MEASure:DMM? 
(Query) 
To the question MEAS:DMM? <INT1|2|3|4> the instrument returns the value of the main 
measurement for the selected channel. 
INT1 to INT4 index are associated with channels 1 to 4. Use the index to find INT5 power 
measurement.  
Before using the command MEAS: DMM? INT5, the instrument must be configured to 
measure the power measurement (see [SENSe]: Function).  
Response format :   <mesure> <NL>  
value format <nrf> 
MEASure:FALL:OVERshoot?
(Query) 
To the question MEAS:FALL:OVER? <INT{1|2|3|4}> the instrument returns the negative 
overshoot of the selected signal. 
Response format:   <measured value><NL> 
value in format <NR2> expressed in percent. 
MEASure:FALL:TIME? 
or 
MEASure:FTIME? 
(Query) 
To the question MEAS:FALL:TIME? <INT{1|2|3|4}> the instrument returns the fall time of the 
selected signal. 
Response format:   <measured value><NL> 
value in format <NR3> expressed in second. 
MEASure:FREQuency? 
(Query) 
To the question MEAS:FREQ? <INT{1|2|3|4}> the instrument returns the frequency of the 
selected signal. 
Response format:   <measured value><NL> 
value in format <NR3> expressed in hertz. 
MEASure:HIGH? 
(Query) 
To the question MEAS:HIGH? <INT{1|2|3|4}> the instrument returns the value of the high level 
level of the selected signal. 
Response format:   <measured value><NL> 
value in format <NR3> expressed in volt. 
MEASure:LOW? 
(Query) 
To the question MEAS:LOW? <INT{1|2|3|4}> the instrument returns the low level value of the 
selected signal. 
Response format:   <measured value><NL> 
value in format <NR3> expressed in volt. 
MEASure:MANual:PHASe? 
(Query) 
To the question MEAS: MAN:PHAS?, the instrument returns the phase of ϕ-cursor in relation 
to cursors 1 and 2. The difference between the cursor 1 and 2 represents 360°. The cursor 1 
equal to 0° and the cursor 2, 360°. 
Response format:   <measured value><NL> 
value in format <NR2> expressed in degree. 
MEASure:MAXimum? 
(Query) 
To the question MEAS:MAX? <INT{1|2|3|4}> the instrument returns the maximum value of the 
selected signal.  
Response format:   <measured value><NL> 
value in format <NR3> expressed in volt. 
MEASure:MINimum? 
(Query) 
To the question MEAS:MIN? <INT{1|2|3|4}> the instrument returns the value minimum of the 
selected signal. 
Response format:   <measured value><NL> 
value in format <NR3> expressed in volt. 