Instructions

Table Of Contents
Fernzugriff
90
MEASure:AC?
(Query)
To the question MEAS:AC? <INT{1|2|3|4}>,<CYCle|INTerval> the instrument returns the
RMS voltage over an integer number of periods (CYCle) or over the measurement interval
(INTerval).
Response format: <measured value><NL>
value in format <NR3> expressed in volt.
MEASure:AMPLitude?
(Query)
To the question MEAS: AMPLitude? <INT{1|2|3|4}> the instrument returns the amplitude of
the selected signal.
MEASure:CURSor:DTIME?
(Query)
To the question MEAS:CURS:DTIME?, the instrument returns the time delay between cursors
1 and 2.
Response format: <measured value><NL>
value in format <NR3> expressed in second.
MEASure:CURSor:DVOLT?
(Query)
To the question MEAS:CURS:DVOLT?, the instrument returns the difference between cursors
1 and 2.
Response format: <measured value><NL>
value in format <NR3> expressed in volt.
MEASure:DMM?
(Query)
To the question MEAS:DMM? <INT1|2|3|4> the instrument returns the value of the main
measurement for the selected channel.
INT1 to INT4 index are associated with channels 1 to 4. Use the index to find INT5 power
measurement.
Before using the command MEAS: DMM? INT5, the instrument must be configured to
measure the power measurement (see [SENSe]: Function).
Response format : <mesure> <NL>
value format <nrf>
MEASure:FALL:OVERshoot?
(Query)
To the question MEAS:FALL:OVER? <INT{1|2|3|4}> the instrument returns the negative
overshoot of the selected signal.
Response format: <measured value><NL>
value in format <NR2> expressed in percent.
MEASure:FALL:TIME?
or
MEASure:FTIME?
(Query)
To the question MEAS:FALL:TIME? <INT{1|2|3|4}> the instrument returns the fall time of the
selected signal.
Response format: <measured value><NL>
value in format <NR3> expressed in second.
MEASure:FREQuency?
(Query)
To the question MEAS:FREQ? <INT{1|2|3|4}> the instrument returns the frequency of the
selected signal.
Response format: <measured value><NL>
value in format <NR3> expressed in hertz.
MEASure:HIGH?
(Query)
To the question MEAS:HIGH? <INT{1|2|3|4}> the instrument returns the value of the high level
level of the selected signal.
Response format: <measured value><NL>
value in format <NR3> expressed in volt.