3. GETTING STARTED
3.1.  Connecting the test leads 
Connect the black lead to the COM socket (for all measurements). Depending on the 
position of the selector switch, connect the red lead as follows: 
Rotary selector switch position  Input terminal 
VLOW Z, VAC+DC, VDC,
ADPDC (ADPAC+DC), Hz,  , Ω,  ,  V Ω
500 mADC (500 mAAC+DC) 500 mA
10 ADC (10 AAC+DC) 10 A
RS232 lead :
Connect the RS232 optical interface (optional kit) to the top of the multimeter casing and 
the DB9F plug to PC (COM port). Refer to § 5. 
3.2.  Switching on the instrument 
The selector switch is on the OFF position. 
Turn the selector switch to the required function. 
All segments of the display come on for a few seconds. The instrument is then ready for 
measurements. 
3.3.  Switching off the instrument 
The instrument can be switched off manually by returning the selector switch to the OFF 
position, or automatically after approximately half an hour if no key is pressed or the switch 
is not operated. 
)
  Note For user safety, automatic shutdown is disabled when a measured 
magnitude (voltage/current) present at the input exceeds dangerous level 
(
 indicator displayed). 
3.4. Special configuration 
To adapt the configuration of the instrument to the measurement environment, the user 
can choose 50 Hz or 60 Hz rejection : 
Switch on with the rotary switch while holding down the MEM key. The selection is 
reversed from the last configuration, is displayed for two seconds and remains backed 
up in non-volatile memory. 
To initialise the RS232 communication, keep the Pk +/- and keys simultaneously 
pressed, then set the rotary switch to the selected position. 
3.5. Multimeter maintenance 
3.5.1. Fuse self-test 
This check can be carried out with the multimeter in use without opening up the casing. 
- 0.63 A fuse (F1) : set the rotary switch to the 
 position and short-circuit the 500 mA and 
VΩ sockets. The display should read approx. 0.001 V. If the digital display indicates an 
overflow « .OL ». The fuse is blown. 
- 10 A fuse (F2) : set the rotary switch to the 
 position and short-circuit the COM and VΩ 
sockets. The display should read approx. 0.001 V. If the digital display indicates an 
overflow « .OL ». The fuse is blown. 