3.5.2. Battery self-test 
When the BAT indication appears on the display, the instrument still has approximately 12 
hours of operation, but specifications can no longer be guaranteed. 
Replace the battery. 
3.5.3.  Replacing the battery or fuses 
* Caution !
Disconnect test leads from any circuit under test, turn the meter off 
and r
emove test leads from the input terminals . 
Use the following procedure: 
1 - Disconnect test leads from any input terminals. 
2 - Using an appropriate tool, slide off the case bottom of the instrument. 
3 - Replace the battery or fuse (respect the value and the type). 
4 - Replace the removable part. 
*  Fitting the fuses :
T
hose fuses are correctly fitted ; those ones are not. 
3.5.4. Cleaning 
Clean the instrument using a damp cloth and soap. Never use abrasive or solvents. 
3.5.5. Storage 
To guarantee the measurement accuracy, after a long period storage in extreme 
environment conditions, wait enough time for the instrument to acclimatise to the working 
environment conditions (see environment specifications). 
Fuses must be replaced 
by fuses of the same 
rating and type.
6LF22 
9 V BATTERY 
F1 0.63 A HBC 18 kA / 600 V  F2 10 A HBC 50 kA / 600 V 