Operation Manual

Portable 4000 count digital multimeter
21
3. FUNCTION DESCRIPTION
3.1
AC or DC voltages
Connect the leads to the multimeter (COM and V terminals)
and
connect it in parallel to the circuit to be tested.
Position the selector sw
itch on the voltmeter function “ V or mV ”.
Automatic range selection : read the measured value.
Memorise it, if necessary, by pressing the “ MEM ” key.
Overload : the beep sounds and “ OL.
blinks.
Specification in AC : from 5 % of the range.
3.1.1 mV
DC
and mV
AC
positions
Range
40 mV
DC
40 mV
AC
Resolution 0.01 mV 0.01 mV
Impedance
1.5 MΩ
Accuracy ± 0.8 % R ± 4 D ± 2 % R ± 6 D
Protection 600 Vrms
Bandwidth 40 Hz to 100 Hz
3.1.2 V
DC
and V
AC
positions
Range 400 mV 4 V 40 V 400 V 600 V
Resolution 0.1 mV 1 mV 10 mV 0.1 V 1 V
Impedance
40 MΩ
>
8 MΩ
Accuracy DC ± 0.3 % R ± 2 D
± 0.5 % R
± 2 D
Accuracy AC ± 1 % R ± 4 D
± 1.5 % R
± 4 D
Protection 600 Vrms
Bandwidth 40-100 Hz 40 - 500 Hz
3.2
Current measurement (AC or DC)
Connect the leads to the multimeter (COM and mA or A terminals)
and connect it in serie to the circuit to be tested.
Position the selector switch on
the appropriated range A, mA or µA.
Read the measured value.
Memorise it, if necessary, by pressing the “ MEM ” key.
Overload : the beep sounds and “ OL.
blinks.
Specification in AC : from 5 % of the range, (
) from 1 A.
Range 400 µA 4 mA 40 mA 400 mA 4 A 10 A ()
Resolution 0.1 µA 0.001 mA 0.01 mA 0.1 mA 1 mA 10 mA
Accuracy
DC
± 1 % R ± 3 D
± 1.5%R
± 7 D
± 1.5%R
± 4 D
Accuracy
AC
± 1.2 % R ± 5 D
± 2%R
± 9 D
± 2.5%R
± 6 D
Voltage
drop
< 0.3 V < 0.7 V < 0.2 V < 0.4 V
Protection fuse 0.63 A HBC fuse 10 A HBC
Bandwidth DC, 40 Hz to 500 Hz
() Acceptance is 20 A overload for 30 s max. with a pause > 5 min.
between each measurement. From 7 A, the measurement is limited to
one minute.