Data Sheet
906211183 - Ed. 2 - 06/2011 - Non contractual document. Confirm specifications before ordering.
State at delivery
Standard versions
- 1 function generator, 1 mains power lead, 1 CD-Rom containing 1 operating manual in
5 languages, 1 programming manual in FR+GB, Labwindows CVI / LabView drivers
Programmable versions
- 1 function generator, 1 mains power lead, 1 CD-Rom containing 1 operating manual
in 5 languages, 1 programming manual in FR+GB, Labwindows CVI / LabView drivers,
1 USB A/B cable
Ethernet version
- As above + 1 cordon Ethernet
To order
- GX305: 5 MHz function generator
- GX310: 10 MHz function generator
- GX310-P: 10 MHz programmable function generator
- GX320: 20 MHz function generator
- GX320-E: 20 MHz programmable function generator
Accessories & spare parts
- AG1066-Z: Set of 2 BNC-Banana leads with rear connection
- HX0106: Set of 2 BNC-BNC leads
- HX0107: Set of 2 BNC-Banana adapters
- HA2004-Z: Set of BNC tee adapters
For assistance and ordering
FRANCE
Chauvin-Arnoux
190, rue Championnet
75876 PARIS Cedex 18
Tel: +33 1 44 85 44 38
Fax: +33 1 46 27 95 59
export@chauvin-arnoux.fr
www.chauvin-arnoux.fr
UNITED KINGDOM
Chauvin Arnoux Ltd
Unit 1 Nelson Ct, Flagship Sq, Shaw Cross Business Pk
Dewsbury, West Yorkshire - WF12 7TH
Tel: +44 1924 460 494
Fax: +44 1924 455 328
info@chauvin-arnoux.co.uk
www.chauvin-arnoux.com
MIDDLE EAST
Chauvin Arnoux Middle East
P.O BOX 60-154
1241 2020 JAL EL DIB (Beyrouth) - LEBANON
Tel: +961 1 890 425
Fax: +961 1 890 424
camie@chauvin-arnoux.com
www.chauvin-arnoux.com
SPECIFICATIONS
GX 305 / GX 310 GX 320
Man-machine interface
Display
LCD (125 x 45 mm) – Adjustable brightness – Display of frequency on 5 digits 20 mm high
Commands on front panel
19 direct-access commands (9 backlit and adjustable) – 1 Main Out On/Off key – 1 digital encoder wheel
Adjustment of signal parameters
Continuous by the encoder, automatic Frequency and Level ranges, selection of the increment digit (F,P,N…)
BNC output terminals on front panel
TTL & Sweep Out outputs TTL, Sweep, Clock and Synch outputs
BNC input terminals on front panel
VCF In input VCG, Gate, Clock and Synch inputs
Continuous signal generation
Frequency
0.001 Hz to 5.000 MHz (10 ranges) (GX305)
0.001 Hz to 10.000 MHz (10 ranges) (GX310)
0.001 Hz to 20.000 MHz (11 ranges)
Resolution / Accuracy
5-digit display – resolution from 1 mHz to 1 kHz according to frequency range
± 20 ppm for F > 10 kHz, ± 30 ppm for F < 10 kHz
Amplitude
1 mV to 20.0 Vpp with open circuit in 3 automatic ranges – 3-digit Vpp or Vrms display
Flatness
< 5 % for 1 mHz< F <10 MHz and ± 1dB up to 20 MHz (GX320) (specs. for level from 0.1 Vpp to 20 Vpp)
Vdc offset
± 10 Vdc with open circuit – accuracy ± 5 % ± 5 mV
Waveforms
Sine / Triangle (max frequency 2 MHz) / Square & "LOGIC" / TTL output
Frequency sweep
Modes
LIN (linear) or LOG (logarithmic)
"INT" internal sweep
"Sawtooth" or "Triangle" mode – Unlimited excursion between "F Start" & "F Stop" (256 steps)
Sweep time adjustable from 10 ms to 100 s
"EXT" external sweep
Sweep by signal < 15 kHz, amplitude ± 10 V –VCF IN input impedance 10 kΩ approx.
Modulations (GX 320)
Internal AM modulation
Modulation by a sine signal with a frequency of 1 kHz
Modulation rate 20 % or 80 %
External AM modulation
Modulation by a signal with a frequency < 15 kHz
Internal FM modulation
Modulation by a sine signal with a frequency of 1 kHz
External FM modulation
Modulation by a signal with a frequency < 15 kHz
SHIFT K function (GX 320)
FSK (Internal/External) = switching between Fstart & F
stop
PSK (Internal/External) = phase switching ± 180°
BURST function
Internal BURST
1 to 65,535 impulsions
Pulse train period from 10 ms to 100 s
External BURST
1 to 65,535 impulsions – Synch/Period by a TTL signal
with a frequency <200 kHz (VCG IN input)
Gate function
Validation of the AC component of "Main Out" by a TTL signal
with a frequency <2 MHz (GATE IN input)
Synch function (GX 320)
Several GX320s in cascade arrangement
Maximum frequency of signals generated 100 kHz
Adjustment of phase shift across ± 180° (resolution 1°)
External frequencymeter
Measurement range
5 Hz to 100 MHz
Accuracy
± 0.05 % + 1 digit
Safety / Max. acceptable voltage
300 V CAT I / 300 V
rms
General specifications
Configuration memories
Storage/Recall of 15 complete instrument configurations
Communication interface
"USB A/B" link for the programmable versions and Ethernet interface (GX 320-E)
Mains power supply
230 V ±10 % (or 115 V ±10 %) – 50/60 Hz – 20 VA max. – Removable lead
Safety / EMC
Safety as per IEC 61010-1 (2001) – EMC as per EN 61326-1 (2004)
Mechanical specifications
227 (L) x 116 (H) x 180 (P) mm – Weight 2.8 kg
Warranty / Origin
3 years – France